Carl Zeiss Axiolab A
3-6 B 40-015 e 06/99
(1) Centering of Objectives (only for Axiolab Pol catalogue No. 450910-0000-000)
Fig. 3-4 Centering of objectives
When the Pol stage (3-4/3) and uncentered objectives are turned, object details above the center of the
eyepiece cross migrate to circular paths (3-4/1 - dashed line). The following procedure is required to
center the objectives:
• Unlock the stage stop by turning screw (3-4/6) by approx. ¼ rotation in counterclockwise direction.
• Swing in transmitted-light objective of medium magnification (20× - 50×) via the nosepiece.
• Insert SW 1.5 key (3-4/2) in the drilled centering holes on the nosepiece.
• Turn the specimen stage and search for the apparent center of rotation (cross (3-4/1)). The center of
rotation is also the center of the stage rotation.
• Use the keys to move this center of rotation to the intersecting point of the eyepiece crosslines and
therefore into the optical axis of the microscope.
• Repeat the procedure with a distinct, small object detail in the direct proximity of the center of the
rotary stage.
• Carefully remove the key.
• Swing in the neighboring objective; search for the center of rotation and the distinct object details
and perform the adjustment procedure as described above.
☞
To retain the centering status, it is absolutely necessary to change the objectives only via the
knurled ring of the objective nosepiece.