Zeiss Axiovert 135M Especificaciones Pagina 48

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 76
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 47
Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć14
2.4.4 VAREL contrast
VAREL contrast provides a reliefĆlike image of specimens and can be used as an alternative to phase
contrast. VAREL contrast can also be used on curved surfaces, e.g. 96 microtiter plates, for which no
contrast can be achieved with phase contrast (no flush ring positioning!).
Preparations as described in section 2.4 and
2.4.1.
Open the aperture diaphragm (2Ć18
ā/ā1)
completely.
Insert the specimen with the suitable holding
frame in the holding clips (2Ć18
ā/ā4) of the
specimen guide.
From the right, push the Phā/āHāVar slide
(2Ć18ā/ā3) into the slider mount on the 0.4 or
0.55 condenser up to the VAREL position.
Swivel in the required VAREL objective.
Use the adjusting screw (2Ć18ā5) to move the
left or right VAREL diaphragm opening (after
selecting the engaging position) until
optimum VAREL contrast is achieved (reliefĆ
like impression).
NOTE Microtiter plates:
Select the opposite VAREL ring of
the diaphragm for illumination at
the edge of the cavity; in the
middle of the cavity, the left or
right VAREL ring can be used.
Figure 2Ć19 VAREL contrast with microtiter plates
Specimen field with relevant pupil image of the VAREL diaphragm
(diaphragm appears rotated by 180°)
12
3
A
B
Figure 2Ć18 VAREL contrast observation
1
2
3
4
5
Vista de pagina 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 75 76

Comentarios a estos manuales

Sin comentarios