Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 3Ć7
Lamp adjustment
CAUTION Never look directly into the
ignited lamp in order to avoid
(irreparable) damage to your
eyes.
Use protective eyewear, e.g. sunĆ
glasses, to protect your eyes when
observing the bright light spot.
• Unscrew an objective and check the image of
the light source via the empty opening on a
piece of paper in the object plane (on the
object stage).
• Use the knurled knob (3Ć5/5) to focus the
collector until both light arcs are in focus.
• Use adjusting screw (3Ć5/1) to set the lamp
axially to the reflector until both light arcs are
equally large, as shown in Figure 3Ć6.
• Use SW 3 screwdriver and the adjusting
screws for height (3Ć5/2) and side (3Ć5/3)
adjustment to centrally position the arc
image next to the arc, as shown in Fig. 3Ć6.
• Screw the objective in the nosepiece again.
• Set the reflector mount to blue excitation,
e.g. by using filter set 09.
• Set the additional filter slider to open light
passage.
Figure 3Ć5 Adjustment of HBO 50
1
2
3
4
5
Figure 3Ć6 Lamp image