Start Capture
To start(/stop) recording - click on the red button icon on
the video toolbar you start your recording as defined in
"Acquisition Setup"
File formats
Images are automatically saved to the library – there is no “Save” option
Individual images can be renamed by clicking on the image name
The original Volocity libraries are stored as "mvd2" files together with a folder containing the image data
in "aisf" files (these cannot be opened). These libraries can be opened either by using Volocity (offline-
licenced server – this is charged by the hour) or with FIJI or ICY (free).
Right click on image to export
Exported TIFF: loses all enhancements but keeps original ("raw") data
- All stacks/ channels are stored in 1 multipage TIFF with the original bit-depth (8 or 12 bit)
Exported TIFF for publication
- Screen version of image, including colours, merge, enhancements, on-screen image resolution,
it is always 8-bit
loss of original ("raw") data!!!
Shutdown procedure
check if anyone is booked after you within 2 hours
If nobody is booked within two hours:
remove specimen
clean oil objective lenses with lens tissue
Important: close incubation chamber completely ( protects from dust)
update time in Sharepoint
secure your data (e.g. copy them to the server)
shut down hardware in reverse order from startup (making sure Fluorescent lamp is switched off even if
you have not used it)
If someone’s booked within two hours:
update usage in Sharepoint
remove your samples
clean objective lenses with fresh lens tissue and close incubation chamber
clear up the desk
save files onto the server or on a mobile hard drive
log off