27 Jan 2015,<qsg-wf1> p. 8/8
Other controls
Auto Exposure
• sets max pixels to 80% (not 100%) (can also be applied to a selected area)
Auto Contrast
• changes the range of pixel values that will be recorded to that of the current image (not good for intensity
measurements
Keypad controls:
• top level button 1 = transmitted light on/off (#2-4 = empty)
• FW button switches between control of:
• FW1 = excitation filter wheel (orbit excitation filters, only used for FRET or CFP+YFP)
• FW2 = emission filter wheel in front of camera, underneath microscope
the orbit excitation filters must be in HOME position for use of the emission filters; if they aren’t,
press the HOME button on the external keyboard
Comentarios a estos manuales