Zeiss LSM 510 META Manual de usuario Pagina 6

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 17
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 5
4. Click the Fast XY button and while the laser is scanning, use the XY Stage Controller and adjust
the fine focus on the microscope body. Clicking "Stop" in the Scan Control window will turn off the
Fast XY continuous scanning (minimize bleaching when you can).
The preliminary image, whether too dim or too bright, usually needs to be optimized. The PMTs need
to be adjusted according to the fluorescence intensity of the sample.
5. Select each channel separately by clicking on the corresponding coloured button.
6. The range indicator display is helpful for setting the gain and offset. Select Palette in the image
window's Select toolbar. From the Color Palette List, select "Range Indicator". In this palette, pixels
with intensity of 0 are blue and pixels with intensity of 255 (for 8 bit images) are displayed red.
7. Start Fast XY and set the Detector Gain until no more than a few % of your stained areas has red
pixels. Move the Amplifier Offset until the background is a mixture of blue and black pixels. (Always
leave the amplifier gain as 1). Make these adjustments for each channel.
Vista de pagina 5
1 2 3 4 5 6 7 8 9 10 11 ... 16 17

Comentarios a estos manuales

Sin comentarios