Zeiss LSM 510 META Manual de usuario Pagina 31

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 66
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 30
Carl Zeiss - Training Application and Support Center
Zeiss LSM 510 META - Guided Tour
31
Minimal Pixel Size determined by Nyquist Sampling
Adjusting the field size (“XY”) to 56 µm with
the 63× lens, would produce a pixel size of
0.1 µm
Field size can be adjusted by changing the objective magnification, or by optical zooming. Changing
from 63 × to 100 × will reduce the field size, but will also reduce the amount of light available.
Brightness of image = Magnification
2
/NA
2
Values are for scan zoom = 1.0
NA
5
10
20
40
63
100
0.15
0.3
0.5
1.3
(oil)
1.4 (
oil
)
1.4 (
oil
)
1.03 µm
0.51 µm
0.31 µm
0.12 µm
0.11 µm
0.11 µm
Pixel sizeMagnification
Vista de pagina 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 65 66

Comentarios a estos manuales

Sin comentarios