Zeiss LSM 510 META Manual de usuario Pagina 39

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 66
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 38
Carl Zeiss - Training Application and Support Center
Zeiss LSM 510 META - Guided Tour
39
Setting up Gain and Offset - Multi Track
1) Select Split XY in the Image window
2) In Palette, select Range indicator
3) Select each channel separately under Channels in the Scan control
window and adjust the Laser intensity, Detector Gain, and Amplifier Offset
as described previously.