Carl Zeiss - Training Application and Support Center
Zeiss LSM 510 META - Guided Tour
64
Shut Down Procedure
1. Go to: Acquire in the LSM menu - Laser – and deactivate HeNe Lasers by
clicking Off to switch off Lasers
2. Go to File - Exit to leave LSM 510 program
3. Shut down the computer
4. Wait until fan of Argon laser has switched off.
5. Turn off the remote control box
6. Switch off the mercury vapour lamp.
Note: To turn off the Argon laser,
first click on Standby, then reduce
output power to 25%. Select Off.