
Carl Zeiss - Training Application and Support Center
Zeiss LSM 510 META - Guided Tour
40
Line Averaging
2) Select Number for averaging. The
more the better for the signal to noise
ratio (max 16) in this case, each line will
be scanned 4 times. But: Averaging
increases the exposure time of the
sample!!
Averaging improves the image by
increasing the signal : noise ratio
Averaging can be achieved line by line,
or frame by frame
1) Select Line or Frame under Mode in
Scan Average within the Mode panel of
the Scan Control window
Comentarios a estos manuales