Start-Up
Carl Zeiss Attaching epi-fluorescence illuminator Axiostar plus
2-30 B 40-815 e 12/01
1 Fastening ring
2 Locking filter
3 P&C reflector module
4 Orientation groove
5 Excitation filter
6 Mounting sheet
Fig. 2-27 Changing filter set in P&C reflector module
2.11.6 Changing filter set in P&C
reflector module
You can combine and attach the filter sets for
the P&C reflector module as needed. Filter sets
or completely equipped P&C reflector modules
can be ordered from Carl Zeiss.
• Remove P&C reflector module (2-27/3) from
the reflector revolver and place in a secure
position.
• Unscrew fastening ring (2-27/1) with
mounting sheet (2-27/6) from the tool set.
• Turn over reflector module so that the filter
(2-27/2 or 5) falls onto a soft surface.
• Insert the locking filter (emission filter)
(2-27/2) and excitation filter (2-27/5) and
secure with the fastening (2-27/1).
The emission filter and excitation filter will be
labeled with an inscription or arrow on the
perimeter. The arrow indicates the installation
direction for the filter into the reflector module
and should point in (see arrow in Fig. 2-27).
To minimize image shift during multiple
fluorescence procedures, the emission filter can
have an additional marking showing the
position of the wedge angle.
The marking should be aligned to the
orientation groove (2-27/4) when inserting an
emission filter in the reflector module. This
guarantees that the wedge angle of the
emission filter has the same defined position in
the reflector modules used. This compensates
for or minimizes the already slight image shift
when Zeiss filter sets are used between
modules.