Start-Up
Carl Zeiss Attaching epi-fluorescence illuminator Axiostar plus
2-32 B 40-815 e 12/01
1 Module half: Excitation
2 Color splitter
3 Spring frame
4 Module half: Emission
5a Eyelets
5b Holding supports
Fig. 2-30 Changing color splitter
1 Angled edge
2 Angled corner
3 Layered side
Fig. 2-31 Color splitter designation
• Tip the top module half up (Excitation)
(2-30/1) and remove the bottom module
half (Emission) from the holding supports
(2-30/5b).
• Remove color splitter (2-30/2) and spring
frame (2-30/3) from bottom module half.
• Remove old color splitter and carefully place
the new one on the spring frame (2-30/3)
with the reflective side pointing up. Place
both parts together in the bottom module
half. Make sure that the side catch of the
spring frame engages in the correct portion
of the bottom module half.
☞
The reflective (layered) side of the color splitter
has an angled edge or corner (Fig. 2-31).
• Place Excitation module half (2-30/1) onto
Emission module half (2-30/4) (holding
support 2-30/5b and eyelets 2-30/5a should
connect). Hold both halves together and
turn in the installation position.
• Replace slotted screws and tighten.
• Attach label with filter combination on the
side of the module.