Zeiss KF 2 ICS Especificaciones Pagina 78

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 100
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 77
Operation
Carl Zeiss Attaching adapters for video cameras Axiostar plus
3-22 B 40-81 e 12/01
The connecting piece 60 - 44 also allows video
adapters with 44 mm interface to be used with
the phototube of the Axiostar plus with 60 mm
interface.
Video adapter
(Cat. No.)
Suitable for Comments
456140
456140-0000-000
Connecting piece 60 - 44
Microscopes with interface 60 mm and
all video adapters for 44 mm interface.
Connects video adapters for 44 mm
interface to microscopes with 60 mm
interface.
The instructions of the camera manufacturer
must also be observed when operating the
video camera.
Attachment and settings:
Loosen three hexagonal screws and remove
dust cover from the camera tube of the
Axiostar plus
.
Screw video adapter or video zoom adapter
with C-mount thread into the video
camera.
Insert video adapter or video zoom adapter
in ENG 2/3" or ENG 1/2 bayonet of the
video camera and clamp it tight.
Insert premounted unit (video camera with
video adapter or video zoom adapter) in
camera tube of the Axiostar plus, align it
and fix it using the three hexagonal screws.
Insert eyepiece with photo reticle in the
binocular tube and align photo reticle
parallel to the camera.
Swivel out folding prism on the binocular
phototube 45° ICS to direct 100 % of the
light to the port. When using the
phototube 30° ICS, pull out the lever for
light path switching completely (100 %
light to camera port).
Set the required zoom magnification factor
via the wheel of the video zoom adapter.
If required, adjust image brightness on the
monitor by changing the lamp brightness
on the microscope stand.
Vista de pagina 77