Operation
Axiostar plus Packing the mount Carl Zeiss
B 40-815 e 12/01 3-27
3.12 Axiostar plus - Packing the
mount
The portable case for the Axiostar plus can be
used to transport the basic Axiostar plus
equipment (stand with binocular tube 45° ICS,
Abbe condenser, mechanical stage, line cable,
dust cover).
• Before inserting the Axiostar plus into the
case, loosen the knurled screw on the stand
(3-17/9) by turning it by ¼, turn the
binocular tube 45°
ICS (3-17/6) around 180°
(eyepieces pointing backwards) and tighten
the knurled screw again.
• Open the four locks of the case and pull the
case cover (3-17/1) upwards to remove it
from the case floor (3-17/2).
• Insert microscope into the recess in the case
floor.
• Push rubber foam component (3-17/4) over
the condenser (3-17/5) and move the
condenser upwards until it contacts the
rubber foam component.
• Pull the binocular tube (3-17/6) completely
apart and slide the rubber foam component
(3-17/7) between the eyepiece tubes until
stop. It might be necessary to apply some
pressure on the rubber foam component
until the holding flaps (3-17/8) are flush
with the upper stand part.
• Line cable, dust cover and any small
accessories can be placed on the case floor
beside the stand and fixed in position using
the Velcro fasteners (3-17/3).
• Attach the case cover, close the locks with a
snap and, if required, lock them with the
enclosed key.