Zeiss LSM 510 Meta Confocal Microscope Instructions
Copyright 2013 University Instrumentation Center UNH
W123 Parsons Hall 23 Academic Way, Durham, NH 03824
Phone: 603-862-2790 Fax: 603-862-4104
www.unh.edu/research/uic
c. Click Channels.
d. The number of Channels buttons shown is the number of fluors
that will be imaged (3 in this example). Click the button of the
fluor with the longest maximum emission wavelength
e. Click 1 button on Pinhole line.
Note the Optical slice thickness.
f. Click the Channels button of the fluor with the next longest
maximum emission wavelength.
g. Adjust Pinhole slider so that Optical slice thickness for second
fluor is the same as for the first fluor (from step 13e).
h. Repeat steps 13f-g if a third fluor is being imaged.
14. Loading sample slide on microscope stage.
a. Tap Microscope button on microscope’s touch screen.
b. Have 5X or 10X objective lens in place before loading a slide on
the stage. If desired lens is not in place, select it by tapping its
button (when Objectives tab is selected).
Always switch objectives using software (on monitor or touch screen), NEVER switch
objectives by hand.
c. To prevent accidentally hitting an objective lens while loading slide onto stage, take one or both
precautions:
Using joystick, move stage forward so that depression that receives
slide is as far forward from objective lenses
as possible.
Joystick has coarse and fine movements. Toggle
between them by pushing straight down on joystick until
you feel a small click.
Tap Load position
(stage will drop).