Zeiss LSM 510 META Manual de usuario Pagina 6

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 13
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 5
Zeiss LSM 510 Meta Confocal Microscope Instructions
Copyright 2013 University Instrumentation Center UNH
W123 Parsons Hall 23 Academic Way, Durham, NH 03824
Phone: 603-862-2790 Fax: 603-862-4104
www.unh.edu/research/uic
drop, waiting for you to add a drop of immersion fluid to the coverslip. After adding immersion
fluid, tap Done
(if you change your mind, Back will return you to the last
objective lens used).
Only use Zeiss Immersol 518F with 40X and 63X oil-immersion objective lenses and either
distilled water or Zeiss Immersol W (for long-term work) with 40X water-immersion objective
lens.
e. To examine your sample by reflected light (RL) (i.e. exciting with
Hg lamp and observing fluorescence produced), turn TL off using TL
knob or TL button, then tap Reflector tab.
We have 3 filter sets for observing fluorescence through eyepieces,
indicated by buttons under Reflector tab labeled:
DAPI (blue fluorescence)
FITC (green fluorescence)
Rhodam[ine] (red fluorescence)
Tap on filter set button most appropriate to the fluorescence you wish to observe, then tap Open
under RL-Shutter to allow light from Hg lamp to reach slide. Tap Close once you have observed
fluorescence in area of interest and centered it in field of view.
Open RL-shutter for as short a period as possible to reduce photobleaching of your fluor(s).
Buttons under Reflector tab labeled Pos. 1, Pos. 6, and DIC TL are only used with TL (halogen
lamp); DAPI, FITC, and Rhodam[ine] are only used with RL (Hg
lamp). Pos. 1 and Pos. 6 are empty positions (contain no optical
elements), either of which should be selected for regular TL viewing.
DIC TL adds differential interference contrast (DIC) optics to the TL
lightpath to add contrast to otherwise virtually invisible samples.
However, the 5X and 10X objective lenses lack the additional DIC
optical elements needed for true DIC.
16. Collecting laser-excited confocal fluorescence images on monitors.
a. Click Find button in Scan Control window.
An Image display window will open.
b. If you are imaging two or
more fluors (3 in this
example), default is for all
channels to be displayed in a
single, overlapping image.
Click Split xy button to see
each channel in a separate
panel.
Fluorescence is first displayed
just for whichever channel
happened to be selected when
the
Find button was clicked. Click a different channel button,
Vista de pagina 5
1 2 3 4 5 6 7 8 9 10 11 12 13

Comentarios a estos manuales

Sin comentarios