8.5 The auxiliary lens for transmitted light
The button to control the auxiliary lens is right above the contrasting method indicator (number 2 in
Figure 13). The lens should be used when imaging with all but the 5x objective. If the auxiliary lens
is in the wrong position, it is impossible to do the Köhler illumination (Chapter 8.6).
8.6 Optimal settings for transmitted light imaging (Köhler illumination)
To achieve optimal image quality some adjustments should be made before imaging. Köhler
illumination needs to be adjusted in the beginning of every transmitted light imaging session. The
elevation and the centering of the condenser can be optimized as follows (Figure 14, the numbers in
the list refer to the arrows in the image):
Select a 10x objective
Select condenser position I/H
Focus on the specimen
Bring the condenser to the uppermost position (2)
Close luminous field diaphragm (1) until the small hole can be seen in oculars
Focus the diaphragm image by moving the condenser (2)
Use centering screws (3) to move diaphragm to the center
Open luminous field diaphragm (1) until its edges disappear from the field of view
Figure 14. Adjusters for field diaphragm (1), condenser elevation (2) and condenser centering
(3).