Zeiss Axioplan 2 imaging and Axiophot 2 Guía de usuario Pagina 5

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 28
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 4
2
1.7 Sample preparation
1.7.1 Selection of fluorochromes
If you want to achieve the best possible resolution, use a fluorochrome whose emission spectrum is
covering as short wavelengths as possible. The resolution of the microscope is determined by the
numerical aperture (NA) of the objective used and the wavelength of the light detected. The larger
the NA, and the shorter the wavelength, the better the resolution.
For thick tissues, image acquisition deeper in the tissue is often a problem because of light
scattering in the tissue. The longer the wavelength of the light, the less it scatters, and the better its
penetration into tissues. Thus, use fluorochromes whose emission spectra are in the far-red region if
you need to image deep inside tissue.
For multilabel imaging, select fluorochromes that are spectrally wide apart. E.g. Alexa488 and
Alexa594 usually have cross-talk between channels if imaged simultaneously, so they should be
imaged sequentially.
1.7.2 Mounting samples
You should either use a mounting medium that will harden or, in the case of non-hardening one,
seal the edges of the cover glass with nail polish or some other sealant. Let the sealant dry well
before you come to the microscope - nail polish is one of the most harmful substances for optics.
Also, unsettled mounting medium, or one containing too much water (slides not dry enough when
mounted), or too much mounting medium, can affect imaging by causing shadows and/or stripes in
the image.
If you are using oil immersion objectives, the refractive index of the mounting medium used should
be close to the refractive index of the immersion oil (1.518). For more information on mounting
media and antifade reagents, visit: http://www.uhnresearch.ca/facilities/wcif/PDF/Mountants.pdf.
Most of the objectives are designed for coverslip thickness 0.17 mm, which corresponds to number
1.5 coverslips.
1.8 Image processing and analysis
MIU has two Imaging Workstations that have software for further image processing and analysis,
such as deconvolution, cell counting, and more. Visit the MIU web site or contact MIU for more
information. Use of Imaging Workstations is free of charge, and MIU staff gives training and help
with the software. MIU also provides customized image analysis tools in case the commercial and
free software do not satisfy your image analysis needs.
2 Turning on the equipment
2.1 Zeiss Axioplan 2 imaging microscope
The main power switch of the microscope (Figure 1) is the green switch located on the right side of
the microscope.
Vista de pagina 4
1 2 3 4 5 6 7 8 9 10 ... 27 28

Comentarios a estos manuales

Sin comentarios