By clicking MEASURE (Figure 24) the software estimates the suitable time of exposure and the
preview live window pops up (Figure 25). Adjust the time of exposure manually if needed and
choose OK. Note that exposure adjustment works slightly different from the live window the
image will not get darker if you shorten the exposure time, but it will be noisier. The image
intensities are multiplied to fit the monitor - hence the noise.
Figure 24. Exposure time setting
Figure 25. Settings for time of exposure.
Add more channels by clicking DUPLICATE (Figure 23). Channels can be acquired either one
channel at the time using SNAP button, or automatically acquiring all channels by clicking START
button at the bottom of the window. Channels can be enabled and disabled by clicking the channel
number with the right mouse button.
10.9 Z-stack imaging
Z-stacks can be acquired by choosing MULTIDIMENSIONAL ACQUISITION/Z. It is possible to
define either the center point of the stack or start and end points. The number of slices can be
entered directly or by specifying the slice distance. OPTIMAL DISTANCE button calculates the
correct distance according to the Nyquist theory. This is important, if you are going process your
stack with 3D-deconvolution software. Note that the correct dye should be selected in
MULTIDIMENSIONAL ACQUISITION/C so the optimal distance could be calculated correctly.