CARE, FUSE REPLACEMENT AND SERVICE
Axio Lab.A1 Instrument care Carl Zeiss
04/2013 430037-7144-001 107
5 CARE, FUSE REPLACEMENT AND SERVICE
5.1 Instrument care
The only care required for the Axio Lab.A1 is as follows:
x Switch the device off each time after use and apply the protective cover (protects against dust and
moisture).
x Do not set the instrument up in a moist environment (max. humidity < 75 %).
x Cover all open tubes with dust caps.
x Remove dust and loose dirt on visible, optical surfaces by means of a fine brush, blower brush, cotton
wool bud, optical paper or cotton cloth.
x Remove water-soluble dirt (coffee, cola, etc.) by breathing on it and wiping with a dust-free cotton
cloth or a moistened cloth. A mild cleaning agent may be added to the water.
x Remove stubborn, oily or greasy dirt (immersion oils, finger prints) using cotton wool buds or a dust-
free cotton cloth and optical cleanser L.
This cleaning agent is manufactured from 85 vol% gasoline and 15 vol% isopropyl alcohol (IPA). The
components are also known by the following names:
Gasoline: rubbing alcohol, light petroleum,
Isopropanol: 2-propyl alcohol,
dimethylcarbinol,
2-hydroxypropane
Clean the optical surfaces with circular movements from the middle to the edge. Exert only slight
pressure on optical surfaces.
The front lens of the condenser Pol should not be cleaned with acetone.
Please observe the following guidelines if the microscope is to be used in a hot and humid climate:
x Keep the instrument in bright, dry and well-aired rooms; humidity < 75 %; especially sensitive parts
such as objectives and eyepieces should be kept in special dry closets.
Precision optical instruments are always susceptible to mould if they are kept and used under following
conditions:
Relative humidity > 75 % over periods exceeding three days at temperatures between +15 °C and
+35 °C.
Set up in dark rooms without sufficient ventilation.
Dust and finger prints on optical surfaces.