Axioskop 40 INSTRUMENT DESCRIPTION
Axioskop 40 FL Technical data Carl Zeiss
B 40-810 e 12/01 1-13
1.7 Technical data
Dimensions (width x depth x height)
Axioskop 40 stand........................................................................ approx. 280 mm x 560 mm x 505 mm
Axioskop 40 FL stand with HBO 100
and adjustable ergo-tube 20°/23 in highest setting........................ approx. 280 mm x 730 mm x 555 mm
Weight
Axioskop 40 ...................................................................................................................... approx. 15 kg
Axioskop 40 FL with HBO 100......................................................................................... approx. 19.5 kg
Ambient conditions
Storage and transport (in packaging)
Permissible ambient temperature ...................................................................................... -10 to + 50 °C
Permissible relative humidity ............................................................................. 20 % to 75 % at +35 °C
Atmospheric pressure ............................................................................................. 800 hPa to 1060 hPa
Operation
Permissible ambient temperature ..................................................................................... +10 to + 35 °C
Permissible relative humidity ............................................................................. 20 % to 75 % at +35 °C
Atmospheric pressure ............................................................................................. 800 hPa to 1060 hPa
Altitude.............................................................................................................................. max. 2000 m
Pollution degree ................................................................................................................................... 2
Operating data
Operating environment.......................................................................................................... clean room
Protection class.......................................................................................................................................I
Protection type................................................................................................................................IP 20
Electrical safety............................................................. in compliance with DIN EN 61010-1 (IEC 1010-1)
including CSA and UL directives
Overvoltage category.............................................................................................................................II
Radio interference suppression....................................................... in accordance with EN 55011, Class B
Insensitivity to interference ......................................................................in accordance with EN 50082-2
Line voltage.............................................................................................................100 to 240 V ±10 %
.......................................................................................................... Change of voltage is not required!
Line frequency...................................................................................................................... 50 to 60 Hz
Power consumption of Axioskop 40 / Axioskop 40 FL ................................................................... 100 VA