START-UP Axioskop 40
Carl Zeiss Initial start-up Axioskop 40 FL
2-4 B 40-810 e 12/01
2 START-UP
The Axioskop 40 / Axioskop 40 FL can be installed, converted and started up by the customer.
However, it is also possible to have the microscope installed or converted by Zeiss service staff for an
extra charge.
Before installing and starting-up the microscope, be sure to carefully read the notes on
instrument safety.
2.1 Initial start-up
2.1.1 Unpacking and installing the microscope
The basic instrument is supplied in a polyethylene case with cardboard packaging.
It contains the following components: Stand with 12 V, 35 W lamp. The binocular tube, objectives,
eyepieces, condenser, microscope mat and various small parts, such as spare lamp, dust cover and tool
bag and other optional accessories are delivered in a separate package.
The following components are factory-installed to the microscope stand: Mechanical stage, specimen
holder, filter mount or overview adapter and, for the Axioskop 40 FL, epi-fluorescence equipment,
adjustment guide (if ordered) and P&C reflector modules.
Remove all components from the packaging
and use the delivery form to check for
completeness.
Place instrument (2-1/1) onto the microscope
mat (2-1/2) on a vibration-free, flat worktable.
The area required by the Axioskop 40
microscope mat is some 500 mm x 400 mm
(width x depth).
Dispose of original packaging properly, or keep
it for storage or for returning the instrument to
the manufacturer.
Fig. 2-1 Setting up the microscope