START-UP Axioskop 40
Carl Zeiss Attachments and conversions Axioskop 40 FL
2-24 B 40-810 e 12/01
Use the SW 3 ball-headed screwdriver to undo
clamping screw (2-29/2) and remove HBO 103
illuminator (2-29/3) from the incident-light tube
(2-29/1) on the microscope stand.
Use focusing knob (2-30/3) to bring the
collector to the frontmost position (in the
direction of the light).
Use a ball-headed screwdriver to undo the
clamping screw for lamp housing (2-30/2) and
carefully remove lamp housing (2-30/1)
upwards from the lamp holder (2-31/3).
Press down the spring lever (2-31/4) and
remove the spent lamp (and transport locking
pin) with the heat sink from holder (2-31/2) in
the lamp holder (2-31/2). There is no need to
remove the wire connection.
Undo fixing screw (2-31/5) on heat sink
(2-31/6) and pull off the heat sink. If the heat
sink does not come off (or cannot be reset),
undo the fixing screw completely, screw it into
the next hole and press it down on the heat
sink.
Dispose of the spent lamp properly.
Insert the new lamp with the smaller diameter
first into the heat sink as far as it will go.
Pay attention to the correct hole
diameter when inserting the lamp
and do not apply any force.
The holder in the heat sink is marked
with H (for HBO) or X (for XBO).
Fig. 2-29 Removing HBO 103 illuminator
Fig. 2-30 Removing HBO 103 housing