START-UP Axioskop 40
Carl Zeiss Initial start-up Axioskop 40 FL
2-16 B 40-810 e 12/01
Fine adjustment without adjustment guide
• Attach the HBO 103 illuminator to the
microscope stand as described in Section 2.1.12
(2).
• Remove one objective from the nosepiece and
switch the empty opening into the function
position.
• Place a white sheet of paper onto the
mechanical stage.
• Completely open the pre-centered luminous-
field and aperture diaphragms in the incident-
light illumination beam path.
• The pupil image with the two focal spot images
is now visible on the white sheet of paper.
When required, reduce the light intensity using
attenuation filters in the 3-position or
6-position filter slider.
• Use knurled knob (2-17/6) to focus on the brighter focal spot via the collector adjustment.
• Use the SW 3 ball-headed screwdriver on adjusting screw (2-17/4) to focus on the slightly darker,
reflected focal spot of the lamp.
• If required, set knurled knob (2-17/6) and adjusting screw (2-17/4) repeatedly until both focal spots
have the same dimension or size.
• Use adjusting screws (2-17/1 and 3) to set the brighter focal spot in accordance with Fig. 2-20.
• Use adjusting screws (2-17/2 and 5) to set the darker focal spot in accordance with Fig. 2-20.
Fig. 2-20 Focal spot image without
adjustment guide on mechanical
stage