OPERATION Axioskop 40
Carl Zeiss Axioskop 40 operation and function controls Axioskop 40 FL
3-6 B 40-810 e 12/01
Color temperature key 3200 K (3-1/7) (3-1/7)
− With the conversion filter switched on and full lamp voltage (12 V), the color temperature comes to
3200 K. This is required for photomicrography using color films (artificial light).
− Lever in front position: conversion filter swiveled out
− Lever in back position: conversion filter swiveled in
Drive knob for movement of the mechanical stage in the y direction (3-1/8)
− Movement of the mechanical stage in the y-direction, maximum travel range: 50 mm
− Adjustable torque (smoothness)
Drive knob for movement of the mechanical stage in the x direction (3-1/9)
− Movement of the mechanical stage in the x direction, maximum travel range: 75 mm
− Adjustable torque (smoothness)
Travel range adjustment for the mechanical stage with torque
− Distance adjustment of the mechanical stage, maximum adjustment range 15 mm
3 levers to swivel filter in / out in the transmitted-light filter magazine (3-1/10)
− 3 switchable neutral filters to adjust image brightness in the field of view,
maximum attenuation factor: 1:4400;
Lever corresponds to (on the stand from front to back): 1.5 %, 6 % and 25 % transmission
− lever in the front position: filter swiveled out
− lever in the back position: filter swiveled in
Centering screw for condenser (3-1/11)
− Centering screws for condenser centering attached on either side of the condenser carrier.