(3) Adjusting the HBO 103 illuminator
Adjustment of the HBO 103 illuminator described
below is performed in two steps: Rough
adjustment and subsequently fine adjustment. The
latter can be performed with or without the
adjustment guide (2-16/2) on the incident-light
tube (2-16/1).
☞
The adjustment guide must be ordered
separately.
Rough adjustment
• Remove the HBO 103 illuminator from the
microscope stand by undoing clamping screw
(2-14/2).
• Switch on the HBO 103 illuminator via the HBO
100 W transformer (2-11/1) and allow it to
warm up to the operating temperature. Take
care to ensure you do not look into the
illuminator’s light-output opening.
• Direct light beam against a projection surface
(wall) at least 3 m away. The focal spot should
correspond roughly to the image in Fig. 2-18.
• Use the knurled knob for collector adjustment
(2-17/6) to focus on the brighter focal spot.
• Use adjusting screw (2-17/4) and the SW 3 ball-
headed screwdriver to locate and focus on the
slightly darker, reflected focal spot.
• Use adjusting screws (2-17/1 and 3) and the
SW 3 ball-headed screwdriver to set the height
and side position of the brighter focal spot in
the same manner as the focal spot image (refer
to Fig. 2-19).
Fig. 2-16 Adjustment guide
Fig. 2-17 Adjusting the HBO 103