START-UP Axioskop 40
Carl Zeiss Initial start-up Axioskop 40 FL
2-12 B 40-810 e 12/01
2.1.11 Inserting and replacing the 12 V,
35 W halogen lamp
• Remove the power cable from the microscope.
• Turn off the microscope and allow it to cool
down for some 15 minutes.
• Undo the knurled screw (2-12/2) and pull up
the lamp housing (2-12/1).
• Hold the lamp carrier (2-13/3) by the knob
(2-12/3), pull it forward and move guide axis
(2-13/4) into the top click-stop position.
• Swivel lamp carrier (2-13/3) backwards, so you
have access to the lamp holder (2-13/2).
• Remove the used lamp (when replacing it) by
pulling it up out of the lamp holder (2-13/2)
without tilting it.
• Push new lamp (2-13/1) with its two socket
pins carefully and without force into the lamp
holder (2-13/2). Do not touch the front
protective glass on the lamp. The lamp must
not be tilted, otherwise the socket pins may be
bent or the lamp holder damaged.
• Swivel the lamp carrier forward again (2-13/3)
until it makes contact.
• Then pull lamp carrier (2-13/3) from the top
click-stop position and move it to the bottom
position. Take care to ensure that the guide axis
is not tilted so that it does not become
jammed. The springs (2-13/5) will center the
lamp.
• Wiggle the knob slightly to ensure that the
lamp is lodged exactly in the compartment
funnel (2-13/6).
• Replace lamp housing (2-12/1) and lock with
knurled screw (2-12/2).
• Reconnect power cable.
Fig. 2-12 Removing the lamp housing
Fig. 2-13 Inserting the 12 V, 35 W halogen
lamp