OPERATION
Image Optimization LSM 510
4-94 B 40-051 e 07/98
Multiple-channel
Requirements:
The suitable lasers are on.
The specimen has been positioned and focused
for scanning.
The slide rod on the microscope tube has been
pulled out as far as it will go.
In the "Acquire" main menu item, click on the
<Config> button.
This opens the "Configuration Control"
window.
Select the necessary configuration by clicking
the <Recording Configurations> button,
selecting a proper configuration and pressing
the <Apply> button (e.g. 488 nm/543 nm).
The configuration loaded is displayed on the
Beam Path and Channel Assignment
panel.
Close the "Recording Configurations" window.
The laser intensity set can be
subsequently optimized for the job via
the <Excitation> button.
Fig. 4-103