INTRODUCTION TO LASER SCANNING MICROSCOPY
Functions LSM 510
5-30 B 40-051 e 07/98
Output defines the name of the result sequence. It will get only those channels which are chosen by the
Input parameter. The buttons labeled with 8 and 12 define the grey value (intensity) resolution in bit.
Normally the result will get the same resolution as the Input sequence. A change will be needed if image
sequences with different resolutions should be combined. Rising the grey value range to 12 bit will not
enhance the display quality or measurement accuracy. The smooth and morphology functions will
produce results with finer gradations.
The Output histogram shows the resulting histogram. They are not editable. The horizontal axis
represents the grey values from 0 to the maximum, which is either 255 or 4095, depending whether the
input is 8 bit or 12 bit. The vertical axis represents the pixel count. The vertical scale of the histogram is
set using the scroll bar. The units are percentages of the grey value distribution maximum. This setting
has no influence on the function.
Parameters:
Input Input image sequence
Output Output image sequence
Threshold Exclusion value - 0...1000
Input L Lower boundary of grey value range Input
Input C Center of grey value range Input
Input H Upper boundary of grey value range Input
Output L Lower boundary of grey value range Output
Output C Center of grey value range Output
Output H Upper boundary of grey value range Output