OPERATION
Annex LSM 510
4-184 B 40-051 e 07/98
<Zero>
Zeros the Current Position display and thus sets the currently set stage position to 0 in relation to X and
Y. The already marked object areas thus receive new X and Y-coordinates.
<Current Position>
Current Position displays the currently set stage position in relation to the zero position.
<Selected Position>
Selected Position displays the coordinates of the position marked on the specimen carrier using the
mouse.
<Marks>
table
Clicking on the arrow keys displays the table of the session-related marked specimen areas. The table
includes the ordinal number, the X-position and the Y-position.
<Move to>
Clicking on the <Move to> button activates the position selected before from the table.
<Remove>
The <Remove> command enables a selected position to be deleted from the table. The position then
also disappears from the specimen carrier display.
CAUTION
: The selected position is deleted, the position with the next number in sequence moves up
one number.
<Remove all>
The <
Remove all
> command deletes all the entries marked in the current session.
Note:
The smallest step width displayed is: 0.1 µm
The smallest step width which can be set is: 0.25 µm
The smallest recommended step width is : 1.0 µm