OPERATION
Image Acquisition (Acquire) LSM 510
4-28 B 40-051 e 07/98
4.5.1 Laser settings
Click on the <Laser> button.
This opens the "Laser Control" window, which shows all lasers connected to the system.
The Laser Selection and Power Status panel,
shows the types, operating statuses and excitation
wavelengths of the lasers available.
Click on the desired laser on the Laser
Selection and Power Status panel.
This highlights the selected laser.
On the "Laser Control" window, activate the laser as follows:
This applies to Coherent UV-Laser 653 II (Enterprise) and Ar-multiline Laser:
Click on the <Standby> button.
Wait for the laser to heat up, until the Status ready - Standby message appears.
Click on the <On> button.
Status ready - On appears.
Use the Power [%] slider to set the laser power which is ideal for the measurement job. To do this,
click on the slider and drag it while keeping the left mouse button depressed.
To change the laser power in steps of 1%, keep the <Ctrl> key depressed while clicking the mouse
on the
or arrow buttons.
Thus, the laser needed for image acquisition is available.
Enterprise: Set power between 50 and 100 % of the maximum tube current.
Optimum operation is at 20 A (Tube Current [A]).
Argon: Set power between 25 and 100 % of the maximum tube current.
Optimum operation is at 8 A.
Fig. 4-31