OPERATION
LSM 510 Quick start
B 40-051 e 07/98 4-11
(2) Move the silver slider on the side of the microscope to the appropriate position - the correct
position will be shown in a message box on the monitor .
(3) Apply appropriate immersion fluid to objective if needed (see attached table for Common
Objective Characteristics).
Select the Objective by clicking in the Objectives area of the Axioplan Control menu or
Axiovert Control menu. Put specimen on the stage - make sure the specimen is mounted
securely and flat.
(4) You can view the specimen in either fluorescence (reflected light) or transmitted light.
(5) To view in fluorescence click in the Reflector Turret area on the
button and click on the
Appropriate FSET (FSET 9
=
FITC, FSET 15
=
Rhodamine, FSET 01
=
DAPI) and set the check
box for Reflected Light On.
(6) To view specimen in transmitted light set the Relflector Turret position to None and check box
for Transmitted Light On.
Fig. 4-11
Dialog unit for
Microscope Axiovert
^^ ^