OPERATION
Shut-Down Procedure LSM 510
4-156 B 40-051 e 07/98
4.11 Shut-Down Procedure
Never shut down the computer by its main switch while your LSM program is still active, or else
you will lose the currently set operating parameters and the images just scanned.
In the "Settings for user" dialog window, which can be activated with the <Options/Settings>
buttons, activate "Laser off on Exit" in the "Shutdown" register. The lasers will then
automatically be switched off when you exit the LSM program.
4.11.1 Exiting the LSM program
Close all open windows of the LSM program by clicking on the closing icon
in the top right corner
of each window.
This closes the respective window and removes the respective icons from the taskbar.
After all dialog windows have been closed, the "LSM 510 Switchboard" window appears.
Fig. 4-166
Click on the <Exit> button.
This terminates the LSM program.
The monitor screen shows the desktop of the Windows NT operating system.